Related Experiment Video
Updated: May 17, 2026

Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
Published on: April 19, 2024
Influence of cycle length variations on antitachycardia pacing effectiveness among ICD patients
Javier Jiménez-Candil1, Jesús Hernández, Ana Martín
1Arrhythmias Unit, Cardiology Department, IBSAL-University Hospital, Salamanca, Spain. jimenezcandil@secardiologia.es
Background:
Antitachycardia pacing (ATP) fails to terminate 5% to 25% of ventricular tachycardias (VTs) occurring in implantable cardioverter-defibrillator patients. We speculated that small fluctuations in VT cycle length (CL) may be related to the efficacy of subsequent ATP.
Objective:
The purpose of this study was to determine the relationship between the R-R variations of the last 12 R-R intervals before ATP and the efficacy of the first ATP attempt.
Methods:
We studied 551 VTs (CL 329±35 ms) occurring in 67 patients. We also analyzed the percentage of variation (P-RR), which was calculated by dividing the mean difference between each R-R interval and the next one by the CL (×100), and the acceleration index (AI), which was calculated by dividing the CL of the first 6 R-R intervals by the CL of the next 6.
Results:
The effectiveness of the first ATP therapy was 81%, being higher in VTs with AI<1 (85% vs 64%; P<.001). After classifying the events according to the tertiles of P-RR, ATP efficiency was better in higher values of P-RR (VTs with AI<1): 99% (third tertile) vs 85% (second tertile) vs 76% (first tertile), P<.001; and for VTs with AI≥1: 94% vs 68% vs 42% (P<.001). By logistic regression, P-RR (%; odds ratio 2.37; P<.001), and AI<1 (odds ratio 4.17; P<.001) were found to be independent predictors of successful first ATP attempts.
Conclusion:
Small changes in CL increase the effectiveness of ATP significantly. VTs with lower degrees of R-R fluctuations, especially when the pattern is a progressive CL shortening, are infrequently terminated by ATP.
More Related Videos
Related Concept Videos
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiomyopathy V: Interprofessional Care
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers
Antiarrhythmic Drugs: Class III Agents as Potassium Channel Blockers
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...