Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
Split calibration curve: an approach to avoid repeat analysis of the samples exceeding ULOQ
Sudipta Basu1, Abdul Basit, Selvan Ravindran
1Department of DMPK & Toxicology, Sai Advantium Pharma Ltd, Building 1, Plot 2, Chrysalis Enclave, International Biotech Park, Phase 2, Hinjewadi, Pune, 411057, India. basu.sudipta@gmail.com
This study introduces a split calibration curve method for bioanalysis, simplifying drug concentration determination. This approach avoids sample dilution and repeat analysis, improving efficiency in drug development and toxicokinetic studies.
Area of Science:
- Analytical Chemistry
- Pharmacokinetics
- Bioanalysis
Background:
- Traditional bioanalysis uses narrow-range calibration curves, which are time-consuming and have limitations.
- Current methods often require sample dilution and repeat analysis for accurate drug concentration determination.
Purpose of the Study:
- To describe a novel split calibration curve approach for bioanalysis.
- To demonstrate the avoidance of sample dilution and repeat analysis without compromising data quality.
Main Methods:
- Implementation of a split calibration curve strategy.
- Validation of the method for determining drug concentrations in plasma samples.
Main Results:
- The split calibration curve approach accurately determined drug concentrations over wide dynamic ranges (e.g., 0.6-15,000 ng/ml for dapsone).
- Successfully validated for analyzing dapsone, cyclophosphamide, and glipizide in a single run.
Conclusions:
- The split calibration curve method eliminates the need for repeat sample analysis.
- This technique is particularly beneficial for toxicokinetic studies with broad dose ranges and studies with limited sample volumes.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Calibration Curves: Correlation Coefficient
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Blank Solutions
In some experimental cases, the reagents, solvents, or lab equipment used in...