Accuracy, agreement, and reliability of fetal cardiac measurements using 4-dimensional spatiotemporal image
Nan Wang1, Hong-ning Xie, Ruan Peng
1Department of Ultrasonic Medicine, Fetal Medical Center, First Affiliated Hospital of Sun Yat-sen University, Guangzhou, Guangdong, China.
Objectives:
The purpose of this study was to evaluate the accuracy, agreement, and reliability of 4-dimensional sonography using spatiotemporal image correlation (STIC) in fetal cardiac measurements during the second and third trimesters.
Methods:
Four-dimensional cardiac STIC volumes were acquired from 150 low-risk singleton pregnancies at gestational ages of 18 to 38 weeks. A total of 11 dimensions of the fetal heart and great vessels were analyzed by two different observers. Accuracy was assessed by comparing the measurements acquired by 4-dimensional STIC with those determined by 2-dimensional echocardiography by calculating the mean error and mean percent error and depicted by Bland-Altman plots. Absolute agreement and reliability of STIC-based measurements were evaluated on the basis of an intraclass correlation coefficient (ICC).
Results:
For STIC-based measurements of cardiac dimensions, good accuracy was achieved by the two experienced observers (mean percent error, -3.10% to 0.87% for observer A; -2.49% to -0.33% for observer B), and absolute agreement was almost perfect (ICC, 95% confidence interval, 0.916-0.993 for observer A; 0.957-0.990 for observer B). Interobserver ICCs among the two observers were between 0.909 and 0.990 (95% confidence interval), whereas intraobserver ICCs for observer A were between 0.894 and 0.989, and those for observer B were between 0.893 and 0.985, showing no significant differences in the reliability of SITC with regard to observers.
Conclusions:
Four-dimensional sonography with STIC is a feasible and accurate method for fetal cardiac dimension measurements in the second and third trimesters. It is in good agreement with 2-dimensional echocardiography and can be performed by different observers.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...