A 30 nm-resolution hard X-ray microscope with X-ray fluorescence mapping capability at BSRF
Qingxi Yuan1, Kai Zhang, Youli Hong
1Beijing Synchrotron Radiation Facility, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing, People's Republic of China.
Abstract:
A full-field transmission X-ray microscope (TXM) operating continuously from 5 keV to 12 keV with fluorescence mapping capability has been designed and constructed at the Beijing Synchrotron Radiation Facility, a first-generation synchrotron radiation facility operating at 2.5 GeV. Spatial resolution better than 30 nm has been demonstrated using a Siemens star pattern in both absorption mode and Zernike phase-contrast mode. A scanning-probe mode fluorescence mapping capability integrated with the TXM has been shown to provide 50 p.p.m. sensitivity for trace elements with a spatial resolution (limited by probing beam spot size) of 20 µm. The optics design, testing of spatial resolution and fluorescence sensitivity are presented here, including performance measurement results.