Image quality and radiation dose of pulmonary CT angiography performed using 100 and 120 kVp
Randy Fanous1, Hany Kashani, Laura Jimenez
1Department of Medical Imaging, Toronto General Hospital, University Health Network, 585 University Ave, NCSB-1C562, Toronto, ON M5G 2N2, Canada.
Objective:
The objective of our study was to compare image quality and radiation dose of pulmonary CT angiography (CTA) performed in the same patient cohort using tube potentials of 100 and 120 kVp.
Materials And Methods:
The study group for this retrospective study was 32 patients (22 women, 10 men) with a mean age of 57 years (age range, 28-83 years; body weight < 100 kg). Patients underwent pulmonary CTA studies performed using 120 and 100 kVp while other scanning parameters were kept constant. Two observers measured image signal and image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), and SNR dose and CNR dose. Two additional observers performed qualitative image quality analysis using a 5-point grading scale (5 = excellent).
Results:
The reduction in tube potential caused image signal to increase by 29% (p < 0.0001), image noise to increase by 68% (p < 0.0001), CNR dose to decrease by 0.8% (p = 0.91) and SNR to decrease by 24% (p = 0.0002) and CNR by 20% (p = 0.0019). Radiation dose (dose-length product) was decreased by 37% to 379.26 mGy × cm at 100 kVp from 604.46 mGy × cm at 120 kVp (p < 0.0001). The median pulmonary arteries image quality scores for observers 1 and 2, respectively, were as follows at 100 kVp: main, 5 and 5; lobar, 5 and 4.5; and segmental, 5 and 4. At 120 kVp, the median image quality scores for observers 1 and 2 were as follows: main, 5 and 5; lobar, 5 and 5; segmental, 4 and 4. A Wilcoxon test analysis indicated no significant difference in image quality between the studies (main, p = 0.59; lobar, p = 0.88; segmental, p = 0.79).
Conclusion:
Pulmonary CTA can be performed using a tube potential of 100 kVp in patients who weigh less than 100 kg (220 lb). Reducing the tube potential from 120 to 100 kVp results in a 37% reduction in radiation dose without a significant impact on diagnostic image quality.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies III: Computed Tomography