Quantitative analysis and design of a spray aerosol inhaler. Part 2: improvements in mouthpiece performance
Michael Hindle1, P Worth Longest
11 Department of Pharmaceutics, Virginia Commonwealth University , Richmond, VA 23298.
Background:
The objective of this study was to utilize previously identified critical design attributes for the capillary aerosol generator as a model spray inhaler in order to develop a second-generation device that minimized aerosol drug deposition in the mouthpiece.
Materials And Methods:
Computational fluid dynamics (CFD) predictive analysis of the critical design attributes indicated that turbulence intensity should be reduced and the effective mouthpiece diameter should be increased. Two second-generation inhaler mouthpieces meeting these specifications were manufactured and tested. The first device (Design 1) implemented a larger cross-sectional area in the mouthpiece and streamlined flow, whereas the second device (Design 2) used a perforated mouthpiece wall. An in vitro deposition study was performed to quantify the deposition of drug mass in the mouthpieces and connected induction ports, and the results were compared with the CFD predictions.
Results:
The two second-generation mouthpieces reduced in vitro aerosol deposition from the original value of 7.8% to values of 2.1% (Device 1) and 4.3% (Device 2), without largely altering the induction port deposition. This was achieved by design alterations aimed at reducing turbulence intensity and increasing the effective mouthpiece diameter. CFD model predictions were in good agreement with the in vitro experimental data.
Conclusions:
A second-generation spray inhaler mouthpiece with low drug deposition was developed using a predictive CFD model and in vitro experiments. Applying this quantitative analysis and design methodology to medical devices, which is similar to the Quality by Design paradigm, could provide significant advantages compared with traditional approaches.
More Related Videos
15:04Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Inhaled Medications
Asthma-IV: Diagnostic and Management
Clinical Assessment for Asthma:
This is the first step in diagnosing and managing asthma. It includes: