Jarvik 2000 biventricular assist device conversion from old pin-shaped bearing pumps to new conical bearing pumps
Shunsuke Saito1, Taichi Sakaguchi, Sigeru Miyagawa
1Department of Cardiovascular Surgery, Osaka University Graduate School of Medicine, 2-2 Yamada-oka Suita, Osaka, 565-0871, Japan. shunsaito@surg1.med.osaka-u.ac.jp
Abstract:
A patient with a Jarvik 2000 (Jarvik Heart, Inc., New York, NY) biventricular assist device had massive hemolysis and device insufficiency of the right ventricular assist device (RVAD) after the anticoagulation was temporarily reversed due to cerebral hemorrhage. After the patient's condition was stabilized by bridge use of CentriMag (Levitronix LLC, Waltham, MA) for right ventricle support, the Jarvik 2000 pumps with pin-shaped bearings were exchanged with new conical bearing pumps. Examination of the removed pumps revealed tight thrombus formation around the pin-shaped bearings.
Related Concept Videos
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
IV Infusion to Oral Dosing: Conversion Methods
Hemodialysis I: Introduction