Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
Two-year radiologic assessment of the Trident Peripheral Self-Locking cup using EBRA
Perrico Nunag1, Angela H Deakin, Ezekiel Oburu
1Department of Orthopaedics, Golden Jubilee National Hospital, Agamemnon Street, Clydebank, Dunbartonshire, UK.
Early radiographic instability was observed in most Trident Peripheral Self-Locking cups, but this did not correlate with initial incomplete seating. Functional outcomes at five years remained good, independent of migration patterns.
Area of Science:
- Orthopedic surgery
- Radiology
- Biomedical engineering
Background:
- The Trident Peripheral Self-Locking cup is a widely used acetabular component in hip arthroplasty.
- Early migration and instability can impact long-term implant survival.
- Understanding factors influencing early cup behavior is crucial for improving patient outcomes.
Purpose of the Study:
- To radiologically assess early migration of the Trident Peripheral Self-Locking cup two years post-implantation.
- To determine if incomplete postoperative seating correlates with early instability.
- To evaluate functional outcomes and revision rates at five years.
Main Methods:
- Retrospective analysis of 30 hip arthroplasty cases using the EBRA (Ensemble de Radiographies Biplan) method.
- Radiographic assessment of cup migration and seating at two years post-surgery.
- Correlation analysis between incomplete seating and migration, and evaluation of Oxford Hip Scores.
Main Results:
- Average total migration was 1.5 mm, with 17 cups showing >1 mm migration and 7 showing >2 mm migration.
- Twenty cups demonstrated incomplete seating postoperatively, but no relationship with migration was found (p = 0.86).
- Oxford Hip Scores significantly improved post-surgery (p = 0.001), independent of migration (p = 0.76).
Conclusions:
- The majority of cups exhibited early radiographic instability, unrelated to incomplete initial seating.
- Five-year functional outcomes appear favorable and are not influenced by migration or initial seating.
- No revisions for aseptic loosening were reported at the five-year follow-up.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT