Analyser position for end-tidal carbon dioxide monitoring in a rebreather circuit
Alastair Ineson1, Kaylene Henderson, David Teubner
1Senior registrar, Department of Anaesthesia, Auckland City Hospital.
Introduction:
A diving rebreather currently nearing release incorporates an infra-red CO2 analyser at the end of the exhale hose and uses the expired gas CO2 measurement made at this position to detect hypercapnia. This configuration may allow exhaled anatomic and mouthpiece dead space gas to mix with alveolar gas in the exhale hose thus falsely lowering the CO2 measurement, especially at low tidal volumes.
Methods:
A test circuit was constructed using a typical rebreather mouthpiece and exhale hose connected into an anaesthetic machine breathing loop. True end-tidal PCO2 was measured in gas sampled from the mouth and compared breath-by-breath to the PCO2 measured in gas sampled at the end of the exhale hose. Two subjects each completed 60 breaths at tidal volumes of 500, 750, 1000, 1500 and 2000 ml.
Results:
There was a small (≤ 0.21 kPa) mean difference between true end-tidal CO2 and end-of-hose CO2 at tidal volumes of 1000 ml or more. However, at lower tidal volumes, the mean difference increased and, at 500 ml, it was 1.04 kPa and 0.70 kPa in subjects 1 and 2 respectively.
Conclusion:
Measurement of the peak exhaled PCO2 at the end of a rebreather exhale hose may provide a reasonable estimation of the true end-tidal CO2 at large tidal volumes, but may significantly underestimate the true end-tidal CO2 at low tidal volumes.
More Related Videos
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
08:34Investigation into Deep Breathing through Measurement of Ventilatory Parameters and Observation of Breathing Patterns
Published on: September 16, 2019
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated. Under...
Respiratory Regulation of Acid-Base Balance
When carbon dioxide levels increase in the blood, more H+ and HCO3⁻ are produced, leading to a...