Sampling and Pretreatment of Tooth Enamel Carbonate for Stable Carbon and Oxygen Isotope Analysis
Published on: August 15, 2018
High-precision measurements of 17O/16O and 18O/16O ratios in CO2
1Institute of Earth Sciences, Hebrew University of Jerusalem, Edmond J. Safra Campus Givat Ram, Jerusalem, 91904, Israel. eugenib@huji.ac.il
Rationale:
Measurements of δ(17)O and δ(18)O values of tropospheric CO(2) are of great importance. However, to be useful, such measurements must be an order of magnitude more precise than in current published literature. With this purpose we developed a new method for high-precision mass spectrometric measurements of (17)O/(16)O and (18)O/(16)O ratios in CO(2), which is presented in this study.
Methods:
The method is based on isotopic exchange equilibration between H(2)O and CO(2) in sealed glass ampoules followed by water fluorination to produce O(2). Dual inlet isotope ratio mass spectrometric (IRMS) measurements of the δ(17)O and δ(18)O values of this O(2) allow δ(17)O values of CO(2) to be obtained with very high precision (0.01 to 0.03 ‰). This method requires about 70 µmol of CO(2).
Results:
Measurements of standard CO(2) gas and atmospheric CO(2) yield reproducibility of 0.01 to 0.03 ‰ for both δ(17)O and δ(18)O values, and 5 per meg for (17)O(excess). Fractionation factors (17)α and (18)α were determined in the H(2)O-CO(2) equilibrium at 25 °C as 1.021254 ± 0.00004 and 1.041036 ± 0.00008, respectively, and the ratio ln(17)α/ln(18)α as 0.5229 ± 0.0001.
Conclusions:
The results demonstrate that the new method is the most accurate analytical procedure ever presented for measurements of (17)O(excess) of CO(2) gas. It is suitable for measurements of CO(2) extracted from relatively small samples of air (~5 L), and is thus useful for monitoring the (17)O(excess) of CO(2) on a broad global scale.
Related Concept Videos
Carbon-13 (¹³C) NMR: Overview
Constant Pressure Calorimetry
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Measurement of Air Content in Concrete
The pressure method,...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...