Radiochromic film measurement of spatial uniformity for a laser generated x-ray environment
J H Fisher1, C D Newlander, R Horton
1Fifth Gait Technologies, Inc., Huntsville, Alabama 35803, USA. Jonathan@5thGait.com
Abstract:
An existing x-ray source application (XRSA) test cassette was modified to hold multiple x-ray filter materials followed by two radiochromic film types (FWT-60 and HD-810 Gafchromic® film) to qualitatively characterize the spectral-spatial uniformity over the XRSA sample field of view. Multiple sets of film were examined and nominal set was determined. These initial, qualitative measurements suggest a low-energy regime (E < 3 keV) spatial anisotropy and spatial isotropy at higher energies (E > 3 keV).