Related Experiment Video
Updated: May 17, 2026

Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
Unexpected difficulty during transcatheter device closure of atrial septal defect associated with right aortic arch
Nageswara Rao Koneti1, Abhijeet Shelke, Srinivasa Kumar Arramraj
1Department of Pediatric Cardiology, Care Hospital, The Institute of Medical Sciences, Hyderabad, Andhra Pradesh, India.
Abstract:
Right aortic arch (RAA) associated with isolated atrial septal defect (ASD) is very rare. We report successful closure of ASD associated with RAA using a 26-mm atrial septal occluder in a 30-year-old male patient. The impingement of right descending aorta in RAA caused malposition of the device in the left atrium. Deployment of the device through the right upper pulmonary vein successfully closed the defect. Follow-up evaluation by computerized tomography scan and echocardiogram showed no pulmonary venous obstruction.
Related Concept Videos
Aneurysm III: Interprofessional Care
Aortic Regurgitation I: Introduction
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Aortic Regurgitation III: Medical Management
Cardiac Catheterization II: Right Heart Catheterization
Aneurysm I: Introduction