Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
An in vitro assessment of MRI issues at 3-Tesla for antimicrobial, silver-containing wound dressings
Kirin B Escher1, Frank G Shellock
1Loyola Marymount University, Los Angeles, CA, USA.
Abstract:
Although no reports of adverse events have been published to date, the presence of metallic dressing ingredients may present an magnetic resonance imaging (MRI) safety concern for patients using silver-containing wound dressings. The purpose of this in vitro study was to test magnetic field interactions (ie, translational attraction and torque), heating, artifacts, and conductivity (ie, electrical resistance) when using MRI at 3-Tesla for two (nonborder and border) silver-containing wound dressings. The results indicated the dressings displayed no magnetic field interactions (deflection angle 0˚; no torque), and in each case, MRI-related heating effects were at the same levels as the background temperature increases (ie, <1.8˚C). The dressings created extremely subtle artifacts (one-for-one relationship) on the MR images. With regard to the conductivity assessments, the average resistance values were 20 kOhm and 1.1 kOhm, respectively, for the nonborder and border wound dressings, which were acceptable levels. The findings show the two silver-containing wound dressings tested will not pose hazards or risks to patients and, thus, are considered "MR safe" according to the current labeling terminology used for medical products, and each dressing may be left in place when a patient undergoes an MRI examination. To date, only a hydrofiber silver-containing dressing has been tested for MRI safety. Because of potential variances in material characteristics, MRI test results are specific to the dressings tested and cannot be applied to other products. Future studies to define the level of silver concentration in dressings that may pose a hazard for performing an MRI are warranted.
Insights
Silver-containing wound dressings are safe for magnetic resonance imaging (MRI). This in vitro study found no adverse magnetic interactions, heating, or significant artifacts, deeming them "MR safe" for patient use during MRI scans.
Area of Science:
- Biomedical Engineering
- Materials Science
- Radiology
Background:
- Silver-containing wound dressings are widely used for their antimicrobial properties.
- The presence of metallic components in dressings raises concerns about safety during magnetic resonance imaging (MRI).
- Previous MRI safety data for silver-containing dressings is limited.
Purpose of the Study:
- To evaluate the MRI safety of two silver-containing wound dressings in vitro.
- To assess magnetic field interactions, heating effects, image artifacts, and electrical conductivity at 3-Tesla MRI.
- To determine if these dressings can be safely left in place during MRI examinations.
Main Methods:
- In vitro testing of two silver-containing wound dressings (nonborder and border) using a 3-Tesla MRI scanner.
- Evaluation of translational attraction, torque, temperature changes, MR image artifacts, and electrical resistance.
- Comparison of results against established safety criteria for medical devices.
Main Results:
- No significant magnetic field interactions (deflection angle 0°, no torque) were observed.
- MRI-related heating was minimal (<1.8°C increase), comparable to background levels.
- Dressings produced negligible artifacts on MR images, with acceptable electrical resistance values (20 kOhm and 1.1 kOhm).
Conclusions:
- The tested silver-containing wound dressings are considered "MR safe" and pose no identified hazards for patients undergoing MRI.
- These dressings can remain in place during MRI examinations.
- Further research is needed to assess MRI safety across a wider range of silver-containing dressings and silver concentrations.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...