Performance of a fourth-generation HIV screening assay and an alternative HIV diagnostic testing algorithm
Muazzam Nasrullah1, Laura G Wesolowski, William A Meyer
1Division of HIV/AIDS Prevention, National Center for HIV, Hepatitis, STD and TB Prevention, Centers for Disease Control and Prevention, Atlanta, GA, USA. snasrullah@cdc.gov
AIDS (London, England)
|November 9, 2012
Summary
The alternative HIV testing algorithm using a fourth-generation assay improves early infection detection and identifies HIV-2. This method shows higher sensitivity than the current algorithm but may yield false positives in indeterminate cases.
Area of Science:
- Clinical diagnostics
- Immunology
- Virology
Background:
- Current HIV testing algorithms rely on third-generation immunoassays and western blot, which may have limitations in early detection.
- Advancements in fourth-generation antigen/antibody assays offer potential improvements in HIV screening.
Purpose of the Study:
- To evaluate the performance of a fourth-generation HIV assay.
- To compare the Centers for Disease Control and Prevention's (CDC's) proposed alternative HIV testing algorithm with the current standard algorithm.
Main Methods:
- A convenience sample of over 10,000 specimens was tested using third-generation and fourth-generation immunoassays, differentiation immunoassays, nucleic acid tests (NAT), and western blot.
- Two algorithms were applied: the current (third-generation immunoassay followed by western blot) and the alternative (fourth-generation immunoassay, differentiation immunoassay, and NAT).
Main Results:
- The alternative algorithm demonstrated improved sensitivity in detecting early HIV-1 infections (seroconverters) compared to the current algorithm.
- The fourth-generation assay showed high specificity (99.91%), with a small number of false positives in western blot-indeterminate specimens.
- The alternative algorithm successfully identified HIV-2 infections and had a higher detection rate for seroconverters.
Conclusions:
- The alternative HIV testing algorithm enhances early infection detection and HIV-2 identification.
- While effective, the alternative algorithm requires careful interpretation, especially with western blot-indeterminate specimens, to manage potential false positives.