Evaluation of nanoparticle tracking analysis for total virus particle determination
Related Experiment Video
Updated: May 17, 2026

Improving Reproducibility to Meet Minimal Information for Studies of Extracellular Vesicles 2018 Guidelines in Nanoparticle Tracking Analysis
Published on: November 17, 2021
Petra Kramberger1, Mateja Ciringer, Aleš Štrancar
1BIA Separations, d,o,o, Mirce 21, SI-5270, Ajdovščina, Slovenia. petra.kramberger@monoliths.com
Abstract:
The NanoSight LM10 with Nanoparticle tracking analysis (NTA) software was evaluated for the quantification of latex particles, adenovirus 5, and influenza virus. The inter-day variability was determined by measuring the same sample over several consecutive days and the method's accuracy was demonstrated by using known concentrations of the subject particles. NTA analysis was also used to quantify chromatographic fractions of adenovirus and influenza virus after purification on a CIM monolithic column. NTA results were compared and evaluated against hemagglutination (HA) and end point dilution assay, determining total and infection virus particle number, respectively. The results demonstrated that nanoparticle tracking analysis is a method for fast estimation of virus concentration in different samples. In addition, it can provide a better insight into the sample status, regarding the level of virus aggregation.