Noise-based tube current reduction method with iterative reconstruction for reduction of radiation exposure in
Junlin Shen1, Xiangying Du, Daode Guo
1Department of Radiology, Xuanwu Hospital of Capital Medical University, No. 45, Chang-Chun St., Xicheng District, Beijing 100053, China. shenjunlin8390@163.com
Purpose:
To investigate the potential of noise-based tube current reduction method with iterative reconstruction to reduce radiation exposure while achieving consistent image quality in coronary CT angiography (CCTA).
Materials And Methods:
294 patients underwent CCTA on a 64-detector row CT equipped with iterative reconstruction. 102 patients with fixed tube current were assigned to Group 1, which was used to establish noise-based tube current modulation formulas, where tube current was modulated by the noise of test bolus image. 192 patients with noise-based tube current were randomly assigned to Group 2 and Group 3. Filtered back projection was applied for Group 2 and iterative reconstruction for Group 3. Qualitative image quality was assessed with a 5 point score. Image noise, signal intensity, volume CT dose index, and dose-length product were measured.
Results:
The noise-based tube current modulation formulas were established through regression analysis using image noise measurements in Group 1. Image noise was precisely maintained at the target value of 35.00 HU with small interquartile ranges for Group 2 (34.17-35.08 HU) and Group 3 (34.34-35.03 HU), while it was from 28.41 to 36.49 HU for Group 1. All images in the three groups were acceptable for diagnosis. A relative 14% and 41% reduction in effective dose for Group 2 and Group 3 were observed compared with Group 1.
Conclusion:
Adequate image quality could be maintained at a desired and consistent noise level with overall 14% dose reduction using noise-based tube current reduction method. The use of iterative reconstruction further achieved approximately 40% reduction in effective dose.
More Related Videos
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...