Related Experiment Video
Updated: May 17, 2026

Non-destructive SPE-UPLC-based Quantification of Aflatoxins and Stilbenoid Phytoalexins in Single Peanut (Arachis spp.) Seeds
Published on: April 19, 2024
Sensitive electrochemical determination of luteolin in peanut hulls using multi-walled carbon nanotubes modified
Dongming Zhao1, Xiuhua Zhang, Lijun Feng
1Ministry of Education, Key Laboratory for the Synthesis and Application of Organic Functional Molecules, College of Chemistry and Chemical Engineering, Hubei University, Wuhan 430062, PR China.
Abstract:
A simple and highly sensitive electrochemical method was developed for the determination of trace-level luteolin, based on the multi-walled carbon nanotubes (MWNTs) modified glassy carbon electrode (GCE). The electrochemical behaviours of luteolin indicate that MWNTs modified glassy carbon electrode (MWNTs/GCE) can greatly enhance the electrocatalytic activity towards the redox of luteolin. It leads to a considerable improvement of the redox peak current for luteolin and allows the development of a highly sensitive voltammetric sensor for the determination of luteolin. A series of experimental parameters including the pH of supporting electrolyte, accumulation potential and time were optimised. The results showed that the oxidative peak currents increased linearly with the concentration of luteolin in the range of 2.0×10(-10) to 3.0×10(-9)M, with a detection limit of 6.0×10(-11)M (S/N=3). The analytical performance of this sensor has been evaluated for detection of luteolin in peanut hulls as a real sample.