Adverse tissue reactions in modular exchangeable neck implants: a report of two cases
S Douglas Werner1, James V Bono, Sumon Nandi
1Department of Orthopaedic Surgery, New England Baptist Hospital, Boston, MA 02120, USA.
Abstract:
Dual-modular femoral stems with exchangeable necks theoretically allow optimization of hip joint biomechanics via selective restoration of femoral anteversion, offset, and limb length. A potential disadvantage is the possible generation of metal ions and debris by fretting and crevice corrosion at the additional stem-neck interface. We present 2 cases of early-onset adverse inflammatory tissue reactions as a result of accelerated corrosion at the stem-neck interface of a dual-modular implant, requiring subsequent revision of well-fixed components.