Related Experiment Video
Updated: May 17, 2026

In Vitro Application of a Wireless Sensor in Flexion-Extension Gap Balance of Unicompartmental Knee Arthroplasty
Published on: May 5, 2023
Implant and limb alignment outcomes for conventional and navigated unicompartmental knee arthroplasty
Gustavo A Valenzuela1, Nathan A Jacobson, Derik J Geist
1Department of Orthopaedic Surgery, Wayne State University, Taylor, MI 48180, USA.
Abstract:
Accurate implant positioning and restoration of lower limb alignment are major requirements for successful long-term results in unicompartmental knee arthroplasty (UKA). Alignment accuracy was compared between navigated-UKA (nUKA) and conventional-UKA (cUKA) groups using a retrospective matched case-control study (n=129, 58 nUKA, 71 cUKA). Mechanical axis (MA), hip-knee-ankle angle (HKA°), coronal implant alignment, and tibial implant posterior slope were measured. No statistically significant difference was observed when comparing MA, HKA° or coronal implant alignment (p>0.05). Statistical significance was seen with tibial component posterior slope (p=0.04, nUKA 4.2°, cUKA 2.9°); and between intra-operative navigationally determined HKA° and post-operative whole-leg standing (WLS) film HKA°. Navigation does not significantly improve UKA alignment compared to conventional methods. Further studies are needed to justify the use of this technology in UKA.