Deep-ultraviolet (UV) resonance raman spectroscopy as a tool for quality control of formulated therapeutic proteins
Sergey Arzhantsev1, Vincent Vilker, John Kauffman
1Division of Pharmaceutical Analysis, Center for Drug Evaluation and Research, US Food and Drug Administration, Saint Louis, MO 63101, USA. sergey.arzhantsev@fda.hhs.gov
Abstract:
A deep-ultraviolet (UV) Raman spectrometer with excitation source tunable from 193 to 210 nm has been built and characterized. The dispersion of the spectrometer over the entire range was measured and described theoretically. The relative sensitivity of the spectrometer was estimated using the integrated intensity ratio of two Raman bands of cyclohexane. Resonance Raman spectra of three formulated insulin products were measured and compared. A band-targeted entropy minimization algorithm was applied to the collected spectra for mixture analysis of insulin products. We conclude that it is feasible to develop robust qualitative methods for quality control of protein-based formulated drug using DUVRR spectroscopy.
Related Concept Videos
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
UV–Vis Spectrometers
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
UV–Vis Spectroscopy of Conjugated Systems
One of the factors influencing λmax is the extent of conjugation in the...
IR and UV–Vis Spectroscopy of Carboxylic Acids
However, the stretching absorptions for the C=O bond vary depending on the structure of carboxylic acids. The C=O bond of the free carboxylic acids shows a higher stretching frequency, 1760 cm−1, while H-bonded carboxylic acids (dimers) exhibit stretching absorptions at a lower frequency, 1710 cm−1. The C=O bond of the...