Wavelength anomalies in ultraviolet-visible spectrophotometry
Related Experiment Video
Updated: May 16, 2026

UV-Vis Spectroscopic Characterization of Nanomaterials in Aqueous Media
Published on: October 25, 2021
1Department of Chemistry, Vanderbilt University, Nashville, TN 37235, USA. joel.tellinghuisen@vanderbilt.edu
Abstract:
Spectral scans of atomic line emission sources with a Shimadzu UV2101PC spectrophotometer show that the nominal wavelength depends upon the instrumental slit width, the wavelength sampling interval, and for some slit widths, also on the specified spectral range. The range dependence is manifested as a smoothing that occurs when the range includes >65 sampled wavelengths, and it affects both the wavelength and the line shape. For spectra not subject to this smoothing, the wavelength error looks like a one-sample misassociation of the wavelength and photometric readings. However, the instrument does reliably move to a specified wavelength, independent of the scan parameter settings. These behaviors do not seem to be documented anywhere but have been present in the software for operating this instrument for about two decades.
More Related Videos
Related Concept Videos
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
UV–Vis Spectrometers
UV–Vis Spectrum
The UV–Vis spectrum of a molecule is the plot of its absorbance versus wavelength. The plot is drawn by taking molar absorptivity (ε) or log ε on the y-axis (ordinate)...
UV–Vis Spectroscopy: Molecular Electronic Transitions
UV–Vis Spectroscopy: Beer–Lambert Law
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...