Measurements of the dose delivered during CT exams using AAPM Task Group Report No. 111
Caroline Descamps1, Mercedes Gonzalez, Edgardo Garrigo
1Instituto de Radioterapia – Fundación Marie Curie,1 Córdoba, Argentina. cdescamps@radioncologia-zunino.org
Abstract:
The computed tomography dose index (CTDI) measured with a 10 cm long pencil ionization chamber placed in a 14 cm long PMMA phantom is typically used to evaluate the doses delivered during CT procedure. For the new generation of CT scanners, the efficiency of this methodology is low because it excludes the contribution of radiation scattered beyond the 100 mm range of integration along z. The AAPM TG111 Report proposes a new measurement modality using a small volume ionization chamber positioned in a phantom long enough to establish dose equilibrium at the location of the chamber. In this work, the AAPM report was implemented. The minimum scanning length needed to obtain cumulative dose equilibrium was evaluated. The equilibrium dose was determined and compared to CTDI values informed by the CT scanner, and the dose values were confirmed with TLD measurements. The difference between doses measured with TLD and with the ionization chamber (IC) was below 1% and the repeatability of the measurements' setup was 0.4%. The measurements showed that the scanning lengths needed to reach the cumulated dose equilibrium were 450 mm and 380 mm for the central and peripheral axes, respectively, which justifies the phantom length. For the studied clinical protocols, the doses measured were about 30% higher than those informed by the CT scanner. For the new generation of CT systems with wider longitudinal detector size or cone-beam technology, the current CTDI measurements may no longer be adequate, and the informed CTDI tends to undervalue the dose delivered. It is therefore important to evaluate CT radiation doses following the AAPM TG111 methodology.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET