Video Movement Analysis Using Smartphones (ViMAS): A Pilot Study
Published on: March 14, 2017
Validation of a novel smartphone accelerometer-based knee goniometer.
Matthew Ockendon1, Robin E Gilbert
1Department of Sports Surgery, Robert Jones and Agnes Hunt Orthopaedic Hospital, Gobowen, Oswestry SY10 7AG, United Kingdom.
The Journal of Knee Surgery
|November 15, 2012
Summary
A new smartphone knee goniometer offers reliable measurement of knee flexion, outperforming traditional devices. This innovation aids in assessing knee function after anterior cruciate ligament surgery.
Area of Science:
- Orthopedics
- Biomedical Engineering
- Sports Medicine
Background:
- Loss of full knee extension post-anterior cruciate ligament (ACL) surgery significantly impairs knee function.
- Accurate and reproducible measurement of fixed knee flexion is challenging with conventional methods.
Purpose of the Study:
- To evaluate the interobserver and intraobserver reliability of a novel smartphone accelerometer-based knee goniometer.
- To compare the smartphone goniometer's performance against a conventional long-armed goniometer for assessing fixed knee flexion deformity.
Main Methods:
- Five healthy male volunteers participated in the study.
- Knee flexion angles were measured using both a conventional telescopic-armed goniometer and a smartphone (iPhone 3GS) knee goniometer employing a trigonometric technique based on tibial inclination.
- Bland-Altman analysis and Pearson's correlation were used to assess validity and reliability.
Main Results:
- The smartphone goniometer demonstrated superior reliability: interobserver correlation (r=0.994) and intraobserver correlation (r=0.982) significantly exceeded those of the conventional goniometer (r=0.952 and r=0.927, respectively).
- The smartphone goniometer exhibited higher precision with interobserver reliability of +/- 2.7 degrees and intraobserver reliability of +/- 4.6 degrees, compared to the conventional goniometer's +/- 8.4 and +/- 9.6 degrees.
- Data from both instruments correlated closely (r=0.947) with a minimal mean difference of -0.4 degrees.
Conclusions:
- The smartphone accelerometer-based knee goniometer is a highly reliable tool for measuring subtle knee flexion in a clinical setting.
- This novel device offers a more accurate and reproducible alternative to conventional goniometers for assessing knee flexion deformities, particularly post-ACL surgery.