Related Experiment Video
Updated: May 16, 2026

Antibiotic Efficacy Testing in an Ex vivo Model of Pseudomonas aeruginosa and Staphylococcus aureus Biofilms in the Cystic Fibrosis Lung
Published on: January 22, 2021
Comparison of five methods for oxacillin susceptibility testing of Staphylococcus aureus isolates from cystic
Marcelo J Mimica1, Rozane L Carvalho, Eitan N Berezin
1Division of Microbiology, Department of Pathology, Santa Casa School of Medicine, São Paulo, SP, Brazil. mjmimica@hotmail.com
Abstract:
Methicillin-resistant Staphylococcus aureus (MRSA) are now a worldwide problem. Cystic fibrosis (CF) patients are commonly colonized and infected by MRSA. Accurate oxacillin susceptibility testing is mandatory for the adequate management of these patients. We performed a comparison of the accuracy of different tests in CF isolates, including methicillin-susceptible S. aureus and MRSA with different SCCmec types, and using the mecA gene as the gold-standard. The sensitivity and specificity of oxacillin disc, Etest, and oxacillin agar screening plate were 100%. Sensitivity of the cefoxitin disc was 85% and specificity was 100%. For clinically relevant isolates, laboratories may consider the use of a combination of two phenotypic methods.