Variability of interface pressure produced by ready-to-wear compression stockings
Purpose:
To assess the variability of interface pressure and changes in this pressure over one month time interval under ready-to-use compression stockings.
Methods:
Fifteen healthy volunteers with a broad range of sizes and shapes of the calves were included in final analysis. Each volunteer sequentially used six pairs of stockings daily for one month each. The six pair set consisted of stockings of three compression classes (class 1: 20-30 mmHg, class 2: 30-40 mmHg and class 3: 40-50 mmHg), made of two materials with different stiffnesses. Interface pressure measurements were performed at B1 point using SIGaT(®) tester (Ganzoni-Sigvaris, St Gallen, Switzerland). Interface pressure was measured in supine and standing positions, and during performing 10 tiptoes. Pressure measurements were performed twice (in the morning and after using stockings for eight hours) on the first day of using each pair of stockings, and repeated on the 30th day.
Results:
At the time of the first use the interface pressure was within the range specified by the manufacturer for 160 out of 180 individual stockings. Twenty stockings (11.1%) produced interface pressure which was 5 mmHg or more outside the range of specified compression class. In 16 of these cases it was only one of the pair of stockings from the same box that produced lower than specified pressure. The pressure under stockings did not change significantly during the day. After one month the interface pressure under the class 1 stockings decreased on average by 1.4 ± 4.3 mmHg (P = 0.013). Class 2 and 3 stockings showed minimal pressure changes which were not statistically significant.
Conclusions:
In vivo measurements of interface pressure should be a requirement for clinical studies of compression stockings, and may be reasonable for ensuring appropriate pressure level in clinical practice.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pressure Variation in a Fluid at Rest
When measuring pressure at two different levels within the fluid, the difference in pressure...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Vascular Resistance
The primary determinants of vascular resistance are vessel diameter, blood viscosity, and vessel length. Among these, vessel diameter plays the most significant role due to the fourth power relationship described by...