Time telling devices used in Danish health care are not synchronized
Mikkel Brabrand1, Susanne Hosbond, Dan Brun Petersen
1Department of Medicine, Sydvestjysk Hospital Esbjerg, 6700 Esbjerg, Denmark. mikkel@brabrand.net
Introduction:
Many patients begin their encounter with the health-care services in an ambulance. In some critical patients, it is pivotal that the timing of treatment and events is registered correctly. When patients are transferred from one health care provider to another, there is a risk that the time telling devices used are not synchronized. It has never been examined if this is a problem in Denmark. We performed the present study to examine if time telling devices used in the pre-hospital setting were synchronized with devices used in emergency departments.
Material And Methods:
We used an on-line atomic clock as reference time. The reference time was compared to watches found in the resuscitation rooms at emergency departments at two hospitals in Denmark. Furthermore, we compared the reference time to the watches on the defibrillators in the ambulances at two ambulance stations.
Results:
The watches in the Emergency Department at Sydvestjysk Hospital Esbjerg had a median deviation of minus three minutes. In the Emergency Department at Hospital Lillebælt Kolding, we found a median deviation of minus 30 seconds. The watches in the defibrillators of 11 ambulances had a median deviation of minus 45 seconds. The maximum deviation between two devices was 19 minutes and 5 seconds, and the maximum deviation between a wall-mounted clock in an emergency department and a defibrillator in an ambulance was five minutes and 22 seconds.
Conclusion:
Examining the time telling devices at two Danish emergency departments and 11 ambulances demonstrated that they are not synchronized.
Funding:
not relevant.
Trial Registration:
not relevant. The study was not registered, as it is an observational study.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Barriers to Effective Communication II
Cultural barriers:
Differences in values, beliefs, religion, knowledge, and tradition can significantly impact communication. Awareness of nonverbal cues is critical, especially when conversing with a patient from a different culture. What appears appropriate in one culture may be inappropriate in another.
Semantic barriers:
As a result of their tendency to use...
Errors occurring during blood pressure monitoring
Several factors...
Documentation in Long-Term and Home Healthcare Setting
Long-Term Care Facilities