Background:
The aim of the present study was to evaluate some analytical performances of the ADVIA Centaur analyzer for the quantitative measurement of 25-OH Vitamin D [25(OH)D] in serum.
Methods:
Serum concentrations of 25(OH)D were determined by a new automated chemiluminescence immunoassay method introduced by Siemens and adapted to an ADVIA Centaur analyzer, and compared with HPLC and a commercial chemiluminescence immunoassay (Liaison DiaSorin).
Results:
The assay displayed a low intra-day (CV < 7.0%) and a low inter-day imprecision (CV < 9.0%). The ADVIA Centaur demonstrated a stronger Spearman's correlation (r = 0.921), better agreement (bias = -0.3 ng/mL), and better concordance correlation coefficient of Lin (P(c) = 0.88) and better kappa index (k = 0.92) with HPLC, than the Liaison DiaSorin assay (r = 0.907, bias = 5.9 ng/mL, p(c) = 0.84, k = 0.60) with HPLC. On the other hand, a significant inverse relationship was observed between parathyroid hormone (PTH) and 25(OH)D concentrations.
Conclusions:
The ADVIA Centaur method is reliable for routine 25(OH)D determination in clinical laboratories
Related Concept Videos
Measurement of Bioavailability: Pharmacodynamic Methods
Automated Microbial Diagnostics
Role of Skin in Vitamin D Synthesis
The solar UV B rays (290-315 nm) are absorbed by the skin, and 7-dehydrocholesterol (provitamin D3) photolyzes it to previtamin D3, which undergoes a rapid transformation to vitamin D3(cholecalciferol).
Measurement of Bioavailability: Pharmacokinetic Methods
Therapeutic Drug Monitoring: Drug Analysis Methods