The effect of perchlorate medication on point-of-care testing
Susanne Hopf1, Monika Bäumel, Michael Gruber
1Department of Anesthesiology, University of Regensburg, Regensburg, Germany. Susanne.Hopf@klinik.uni-regensburg.de
Background:
To create awareness for mismeasurements of ion-selective electrodes caused by patients' medications and to prevent severe consequences in an intensive care therapy.
Case Presentation:
A 73-year-old woman presented with severe carbimazol-induced toxic epidermal necrolysis. After replacement of carbimazol by sodiumperchlorate, we detected a huge discrepancy in the measurement of ionised calcium by two different Point-of-Care (POCT) systems. While Siemens Rapidlab 865 showed a severe hypocalcemia at all times, the Radiometer 600 system first presented a severe hypercalcemia and, in the course of the following days, also a hypocalcemia but with significantly (p < 10 to the -45) higher values than the Rapidlab system. Furthermore, the POCT systems detected a pseudohypocalcemia at times where we think there was a normo- or even hypercalcemia which led to incorrect therapy with excessive substitution of ionised calcium.
Conclusions:
The substance sodiumperchlorate, which is well established in Europe for hyperthyreosis therapy, caused malfunctions of analyses of ionised calcium by POCT systems.
More Related Videos
Related Concept Videos
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Precipitation Titration Curve: Analysis
Precipitation Titration: Overview
A precipitation titration curve demonstrates the change in concentration of the titrant or analyte upon adding the...
EDTA: Indirect and Alkalimetric Titration
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides and...