Improved corner detection by ultrasonic testing using phase analysis
Related Experiment Video
Updated: May 16, 2026

Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Patrik Broberg1, Anna Runnemalm, Mikael Sjödahl
1University West, SE-461 86 Trollhattan, Sweden. patrik.broberg@hv.se
Abstract:
In ultrasonic testing, corners are used for sensitivity calibration in the form of notches, for measuring the sound velocity in the material, and as known reference points during testing. A 90° corner will always reflect incoming waves in the opposite direction due to a double reflection and therefore give a strong echo. This article presents a method for separating the echo from a corner from other echoes and more accurately find the position of the corner. The method is based on analysing the phase of the reflected signal. The proposed method was tested on a steel calibration block and the width of the indication was reduced by up to 50% compared to the amplitude signal. This results in a more accurate positioning of the corner. Using the phase instead of the amplitude will also improve the reliability, since reflections other than from corners will disappear.