Filtered back projection, adaptive statistical iterative reconstruction, and a model-based iterative reconstruction
Zsuzsanna Deák1, Jochen M Grimm, Marcus Treitl
1Department for Clinical Radiology, Ludwig-Maximilians-University, Campus Innenstadt, Nussbaumstrasse 20, 80336 Munich, Germany. zsuzsanna.deak@med.uni-muenchen.de
Purpose:
To compare objective and subjective image quality parameters of three image reconstruction algorithms of different generations at routine multidetector computed tomographic (CT) examinations of the abdomen.
Materials And Methods:
This institutional review board-approved study included 22 consecutive patients (mean age, 56.1 years ± 15.8 [standard deviation]; mean weight, 79.1 kg ± 14.8) who underwent routine CT examinations of the abdomen. A low-contrast phantom was used for objective quality control. Raw data sets were reconstructed by using filtered back projection (FPB), adaptive statistical iterative reconstruction (ASIR), and a model-based iterative reconstruction (MBIR). Radiologists used a semiquantitative scale (-3 to +3) to rate subjective image quality and artifacts, comparing both FBP and MBIR images with ASIR images. The Wilcoxon test and the intraclass correlation coefficient were used to evaluate the data. Measurements of objective noise and CT numbers of soft tissue structures were compared with analysis of variance.
Results:
The phantom study revealed an improved detectability of low-contrast targets for MBIR compared with ASIR or FBP. Subjective ratings showed higher image quality for MBIR, with better resolution (median value, 2; range, 1 to 3), lower noise (2; range, 1 to 3), and finer contours (2; range, 1 to 2) compared with ASIR (all P < .001). FBP performed inferiorly (0, range, -2 to 0]; -1 [range, -3 to 0]; 0 [range, -1 to 0], respectively; all, P < .001). Mean interobserver correlation was 0.9 for image perception and 0.7 for artifacts. Objective noise for FBP was 14%-68% higher and for MBIR was 18%-47% lower than that for ASIR (P < .001).
Conclusion:
The MBIR algorithm considerably improved objective and subjective image quality parameters of routine abdominal multidetector CT images compared with those of ASIR and FBP.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...