Stroke volume optimization in elective bowel surgery: a comparison between pulse power wave analysis (LiDCOrapid) and
J Nordström1, C Hällsjö-Sander, R Shore
1Department of Anaesthesiology, Surgical Services and Intensive Care Medicine, Karolinska University Hospital, 171 76 Stockholm, Sweden.
Background:
Goal-directed fluid therapy improves outcome in major surgery. We evaluated a new device (LiDCOrapid) against our standard oesophageal Doppler method (ODM) for stroke volume (SV) optimization during colorectal surgery.
Methods:
This was an observational study in 20 patients undergoing major colorectal surgery within a fast-track protocol. We compared SV values measured simultaneously by LiDCOrapid and ODM before and after 86 fluid challenges. We also evaluated the LiDCOrapid dynamic indices SV variation (SVV) and pulse pressure variation (PPV) as predictors for volume responsiveness, defined as an increase in SV ≥ 10% after 200 ml of colloid.
Results:
SV increased ≥ 10% after 27 out of 86 fluid challenges. For 172 paired SV values, the overall correlation was r=0.39, and bias (limits of agreement) -28 (-91-35) ml, percentage error 70%. The ability of LiDCOrapid to track changes in SV was weak with a concordance rate of 80%, and a sensitivity and specificity of 48% and 81%, respectively, to detect a positive fluid challenge. The area under the curve values (with 95% confidence intervals) for SVV and PPV were 0.72 (0.60-0.83) and 0.66 (0.52-0.79), respectively, indicating low predictive capacity in these setting.
Conclusions:
LiDCOrapid and ODM devices are not interchangeable. We cannot recommend that the LiDCOrapid replace the standard Doppler method until further device-specific outcome studies on volume optimization are available. The dynamic indices SVV and PPV add little value to a fluid optimization protocol, and should not replace SV measurements with a validated technique.
Related Concept Videos
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...