Performance of an automatic quantitative ultrasound analysis of the fetal lung to predict fetal lung maturity
Montse Palacio1, Teresa Cobo, Mònica Martínez-Terrón
1Department of Maternal-Fetal Medicine, Institut Clínic de Ginecologia, Obstetrícia i Neonatologia, Hospital Clinic, Universitat de Barcelona, Barcelona, Spain.
Objective:
The objective of the study was to evaluate the performance of automatic quantitative ultrasound analysis (AQUA) texture extractor to predict fetal lung maturity tests in amniotic fluid.
Study Design:
Singleton pregnancies (24.0-41.0 weeks) undergoing amniocentesis to assess fetal lung maturity (TDx fetal lung maturity assay [FLM]) were included. A manual-delineated box was placed in the lung area of a 4-chamber view of the fetal thorax. AQUA transformed the information into a set of descriptors. Genetic algorithms extracted the most relevant descriptors and then created and validated a model that could distinguish between mature or immature fetal lungs using TDx-FLM as a reference.
Results:
Gestational age at enrollment was (mean [SD]) 32.2 (4.5) weeks. According to the TDx-FLM results, 41 samples were mature and 62 were not. The imaging biomarker based on AQUA presented a sensitivity 95.1%, specificity 85.7%, and an accuracy 90.3% to predict a mature or immature lung.
Conclusion:
Fetal lung ultrasound textures extracted by AQUA provided robust features to predict TDx-FLM results.