[The cementless artificial hip cup--implant components with reliable long-term stability?]
Related Experiment Video
Updated: Oct 8, 2026

Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Abstract:
A hip joint simulator was developed to analyse the mechanism of loosening of cementless artificial hip cups. The machine induces vibrating motions and asymmetrical tilt shock loadings of the artificial cup. On measuring the primary stability of threaded cups, the simulation tests performed on pelvis substitute models, and animal and human acetabula failed to show any loosening of threaded cups fixed in place with a screw-in torque of more than 10 Nm. Instable cups became loose all the sooner, the lower the preload between the cup and femoral head. This demonstrates the importance of both the screw-in torque of the cup and the training status of the periarticular muscles.

