Kinetics of intravenous radiographic contrast medium injections as used on CT: simulation with time delay
1Heilig-Hartziekenhuis Roeselare-Menen, Belgium. comed@skynet.be
Objectives:
To develop a multicompartment model of only essential human body components that predicts the contrast medium concentration vs time curve in a chosen compartment after an intravenous injection. Also to show that the model can be used to time adequately contrast-enhanced CT series.
Methods:
A system of linked time delay instead of ordinary differential equations described the model and was solved with a Matlab program (Matlab v. 6.5; The Mathworks, Inc., Natick, MA). All the injection and physiological parameters were modified to cope with normal or pathological situations. In vivo time-concentration curves from the literature were recalculated to validate the model.
Results:
The recalculated contrast medium time-concentration curves and parameters are given. The results of the statistical analysis of the study findings are expressed as the median prediction error and the median absolute prediction error values for both the time delay and ordinary differential equation systems; these are situated well below the generally accepted maximum 20% limit.
Conclusion:
The presented program correctly predicts the time-concentration curve of an intravenous contrast medium injection and, consequently, allows an individually tailored approach of CT examinations with optimised use of the injected contrast medium volume, as long as time delay instead of ordinary differential equations are used.
Advances In Knowledge:
The presented program offers good preliminary knowledge of the time-contrast medium concentration curve after any intravenous injection, allowing adequate timing of a CT examination, required by the short scan time of present-day scanners. The injected volume of contrast medium can be tailored to the individual patient with no more contrast medium than is strictly needed.
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...
Compartment Models: Two-Compartment Model
Compartment Models: Single-Compartment Model