Error mechanisms of the oscillometric fixed-ratio blood pressure measurement method
Jiankun Liu1, Jin-Oh Hahn, Ramakrishna Mukkamala
1Department of Electrical and Computer Engineering, Michigan State University, 428 S. Shaw Lane, Rm. 2120 Engineering Building, East Lansing, MI 48824-1226, USA.
Abstract:
The oscillometric fixed-ratio method is widely employed for non-invasive measurement of systolic and diastolic pressures (SP and DP) but is heuristic and prone to error. We investigated the accuracy of this method using an established mathematical model of oscillometry. First, to determine which factors materially affect the errors of the method, we applied a thorough parametric sensitivity analysis to the model. Then, to assess the impact of the significant parameters, we examined the errors over a physiologically relevant range of those parameters. The main findings of this model-based error analysis of the fixed-ratio method are that: (1) SP and DP errors drastically increase as the brachial artery stiffens over the zero trans-mural pressure regime; (2) SP and DP become overestimated and underestimated, respectively, as pulse pressure (PP) declines; (3) the impact of PP on SP and DP errors is more obvious as the brachial artery stiffens over the zero trans-mural pressure regime; and (4) SP and DP errors can be as large as 58 mmHg. Our final and main contribution is a comprehensive explanation of the mechanisms for these errors. This study may have important implications when using the fixed-ratio method, particularly in subjects with arterial disease.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.