Locked plate fixation of the comminuted distal fibula: a biomechanical study
Neil J White1, David T Corr, James P Wagg
1Orthopaedic Trauma, the Royal Infirmary of Edinburgh, Scotland, UK. neiljwhite@gmail.com
Background:
The purpose of this study was to compare the biomechanical properties of locked versus nonlocked lateral fibular bridge plating of comminuted, unstable ankle fractures in a mode of catastrophic failure.
Methods:
We created comminuted Weber C fractures in 8 paired limbs from fresh cadavers. Fractures were plated with either standard or locked one-third tubular bridge plating techniques. Specimens were biomechanically evaluated by external rotation to failure while subjected to a compressive load approximating body weight. We measured the angle to failure, torque to failure, energy to failure and construct stiffness.
Results:
There was no significant difference in construct stiffness or other biomechanical properties between locked and standard one-third tubular plating techniques.
Conclusion:
We found no difference in biomechanical properties between locked and standard bridge plating of a comminuted Weber C fibular fracture in a model of catastrophic failure. It is likely that augmentation of fixation with K-wires or transtibial screws provides a construct superior to locked bridge plating alone. Further biomechanical and clinical analysis is required to improve understanding of the role of locked plating in ankle fractures and in osteoporotic bone.