Continuous noninvasive hemoglobin monitor from pulse ox: ready for prime time?
Bellal Joseph1, Pantelis Hadjizacharia, Hassan Aziz
1Department of Surgery, University of Arizona, Tucson, AZ, USA. bjoseph@surgery.arizona.edu
Background:
Advances in technology have allowed for continuous noninvasive hemoglobin monitoring (SpHb), which may enable earlier detection of hemorrhage and more efficient surgical and/or blood transfusion management. The use of SpHb has not been described in the trauma population. The purpose of the present study was to evaluate the accuracy of a SpHb measurement device in severely injured trauma patients.
Methods:
We performed a prospective cohort analysis of severely injured trauma patients admitted to the intensive care unit (ICU) at our level I trauma center over a 6 month period. Serial IHb (invasive hemoglobin) levels and SpHb for the first 72 h were measured. Each SpHb measurement was matched with a corresponding IHb measurement. We defined normal Hgb as >8 mg/dL and low Hgb as <8 mg/dL. Data were then grouped based on Hgb level. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and Spearman correlation coefficient plot were calculated.
Results:
A total of 23 trauma patients with 89 data pairs were reviewed. Eighty-six percent of the patients were male with a mean age of 32 years and a mean injury severity score (ISS) of 21.1 ± 14. Invasive hemoglobin had a range of 7.2-16.9 and SpHb had a range of 3.3-15.2. The average mean and difference between IHb and SpHb were 10.7 and 1, respectively. Continuous noninvasive hemoglobin measurement did not record data points 13.5% of the time. The Spearman correlation plot revealed a correlation of R = 0.670 (p < 0.001). After dichotomization with Hgb > 8, SpHb was found to have a sensitivity of 91%, PPV 96%, specificity 40%, NPV 20%, and an accuracy of 88%.
Conclusions:
The continuous noninvasive hemoglobin monitor does not appear to represent serum hemoglobin levels accurately in severely injured trauma patients. However, we were able to identify utility for this noninvasive tool when Hgb was dichotomized into normal or low levels.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Holter Monitor: 24-Hour Monitoring