Related Experiment Video
Updated: May 16, 2026

Real-time In Vitro Monitoring of Odorant Receptor Activation by an Odorant in the Vapor Phase
Published on: April 23, 2019
Development of an electronic nose for environmental odour monitoring
Licinia Dentoni1, Laura Capelli, Selena Sironi
1Politecnico di Milano, Department of Chemistry, Materials and Chemical Engineering Giulio Natta, 20133 Milano, Italy. ldentoni@chem.polimi.it
Abstract:
Exhaustive odour impact assessment should involve the evaluation of the impact of odours directly on citizens. For this purpose it might be useful to have an instrument capable of continuously monitoring ambient air quality, detecting the presence of odours and also recognizing their provenance. This paper discusses the laboratory and field tests conducted in order to evaluate the performance of a new electronic nose, specifically developed for monitoring environmental odours. The laboratory tests proved the instrument was able to discriminate between the different pure substances being tested, and to estimate the odour concentrations giving correlation indexes (R2) of 0.99 and errors below 15%. Finally, the experimental monitoring tests conducted in the field, allowed us to verify the effectiveness of this electronic nose for the continuous detection of odours in ambient air, proving its stability to variable atmospheric conditions and its capability to detect odour peaks.
Related Concept Videos
Olfaction
The olfactory receptors are embedded in the cilia of the...
Microbial Biosensors
Physiology of Smell and Olfactory Pathway
The olfactory...