A Tuberculosis Molecular Bacterial Load Assay (TB-MBLA)
Published on: April 30, 2020
"Proof-of-concept" evaluation of an automated sputum smear microscopy system for tuberculosis diagnosis
James J Lewis1, Violet N Chihota, Minty van der Meulen
1London School of Hygiene and Tropical Medicine, London, United Kingdom. james.lewis@lshtm.ac.uk
Background:
"TBDx" is an innovative smear microscopy system that automatically loads slides onto a microscope, focuses and digitally captures images and then classifies smears as positive or negative using computerised algorithms.
Objectives:
To determine the diagnostic accuracy of TBDx, using culture as the gold standard, and compare this to a microscopist's diagnostic performance.
Methods:
This study is nested within a cross-sectional study of tuberculosis suspects from South African gold mines. All tuberculosis suspects had one sputum sample collected, which was decontaminated prior to smear microscopy, liquid culture and organism identification. All slides were auramine-stained and then read by both a research microscopist and by TBDx using fluorescence microscopes, classifying slides based on the WHO classification standard of 100 fields of view (FoV) at 400× magnification.
Results:
Of 981 specimens, 269 were culture positive for Mycobacterium tuberculosis (27.4%). TBDx had higher sensitivity than the microscopist (75.8% versus 52.8%, respectively), but markedly lower specificity (43.5% versus 98.6%, respectively). TBDx classified 520/981 smears (53.0%) as scanty positive. Hence, a proposed hybrid software/human approach that combined TBDx examination of all smears with microscopist re-examination of TBDx scanty smears was explored by replacing the "positive" result of slides with 1-9 AFB detected on TBDx with the microscopist's original reading. Compared to using the microscopist's original results for all 981 slides, this hybrid approach resulted in equivalent specificity, a slight reduction in sensitivity from 52.8% to 49.4% (difference of 3.3%; 95% confidence interval: 0.2%, 6.5%), and a reduction in the number of slides to be read by the microscopist by 47.0%.
Discussion:
Compared to a research microscopist, the hybrid software/human approach had similar specificity and positive predictive value, but sensitivity requires further improvement. Automated microscopy has the potential to substantially reduce the number of slides read by microscopists.
Related Concept Videos
Pulmonary Tuberculosis IV
Several diagnostic approaches are used to detect TB. The conventional method is the Tuberculin Skin Test (TST), also known as the Mantoux test. However, this method has...
Sputum Studies I: Gram Stain, cytology, and Acid-fast smear and culture
Gram Stain
The Gram Stain is an integral part of sputum studies. It involves the staining of sputum, which permits...
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...
Automated Microbial Diagnostics
Pulmonary Tuberculosis III
The first classification is based on the development of the disease, and it includes the following categories: