Related Experiment Video
Updated: May 16, 2026

12:18
Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
Three-dimensional shear-wave elastography for differentiating benign and malignant breast lesions: comparison with
Ji Hyun Youk1, Hye Mi Gweon, Eun Ju Son
1Department of Radiology, Gangnam Severance Hospital, Yonsei University College of Medicine, 211, Eonju-ro, Gangnam-Gu, Seoul 135-720, South Korea. jhyouk@yuhs.ac
European Radiology
|December 6, 2012
Summary
Two-dimensional shear-wave elastography (2D SWE) demonstrates superior diagnostic performance for breast lesions compared to 3D SWE. Combining 2D SWE with B-mode ultrasound significantly improves diagnostic accuracy.
Area of Science:
- Medical Imaging
- Diagnostic Ultrasound
- Breast Imaging
Background:
- Shear-wave elastography (SWE) is an ultrasound technique measuring tissue stiffness.
- Evaluating breast lesions with SWE can enhance diagnostic information.
- Comparison between 2D and 3D SWE for breast lesion assessment is crucial.
Purpose of the Study:
- To compare the interobserver agreement of 3D SWE and 2D SWE for breast lesions.
- To evaluate the diagnostic performance of 3D SWE versus 2D SWE.
- To assess the added diagnostic value of SWE to B-mode ultrasound.
Main Methods:
- 163 breast lesions in 146 women were analyzed using B-mode ultrasound and SWE.
- Two radiologists reviewed B-mode, 2D SWE, and 3D SWE datasets.
- Interobserver variability was assessed using kappa statistic; diagnostic performance was evaluated using AUC.
Main Results:
- 2D SWE showed higher agreement and superior diagnostic performance (AUC) compared to 3D SWE.
- The combination of 2D SWE with B-mode ultrasound significantly increased diagnostic performance.
- 3D SWE, when added to B-mode ultrasound, did not yield a significant diagnostic improvement.
Conclusions:
- 3D SWE offers good interobserver agreement but is diagnostically inferior to 2D SWE.
- 2D SWE significantly enhances the diagnostic capabilities of B-mode ultrasound for breast lesions.
- 3D SWE's diagnostic performance is not significantly improved by its combination with B-mode ultrasound.