Comparison of the delay in processing time and protective plastic cases in two phosphor plate systems
Ali Murat Aktan1, Mehmet Ertuğrul Ciftçi, Faruk Akgünlü
1Department of Oral and Maxillofacial Radiology, Gaziantep University, 27310 Gaziantep, Turkey.
Thescientificworldjournal
|December 6, 2012
Summary
Scanning delay degrades phosphor plate system (PSP) image quality. Using manufacturer
Area of Science:
- Dental radiology
- Image processing
- Radiographic technology
Background:
- Phosphor plate systems (PSPs) are widely used in digital dental radiography.
- Optimizing image quality and understanding factors affecting it are crucial for accurate diagnosis.
Purpose of the Study:
- To evaluate the impact of scanning delay on two PSP systems (Dürr Dental and Digora Optime).
- To assess the performance of different protective plastic cases in preserving image quality.
Main Methods:
- Two PSPs were exposed with an aluminum step wedge and placed in three types of protective cases.
- Radiographs were processed at various intervals post-exposure (immediately to 480 minutes).
- Mean gray values (MGVs) of the step wedge were analyzed using ANOVA.
Main Results:
- Significant differences in MGVs were observed between PSPs and across scan delays.
- Image quality generally decreased with increased scanning delay.
- Dürr plates in their original cases showed less degradation compared to other groups.
Conclusions:
- Scanning delay negatively affects PSP image quality.
- Manufacturer's original protective cases offer superior performance in maintaining image quality for PSPs.
Related Concept Videos
Fluorescence and Phosphorescence: Instrumentation
Fluorometers and spectrofluorometers are two types of instruments used for measuring molecular fluorescence. These instruments differ in how they select excitation and emission wavelengths and the type of light sources they utilize. Fluorometers use absorption interference filters to choose excitation and emission wavelengths. The excitation source in a fluorometer is typically a low-pressure mercury vapor lamp that emits intense lines distributed throughout the ultraviolet and visible regions.
Photoluminescence: Applications
Photoluminescence offers a wide range of applications due to its inherent sensitivity and selectivity. This technique allows for both direct and indirect analyses of the analyte. Direct quantitative analysis is possible when the analyte exhibits a favorable quantum yield for fluorescence or phosphorescence. However, an indirect analysis may be feasible if the analyte is not fluorescent or phosphorescent, or if the quantum yield is unfavorable. Indirect methods include reacting the analyte with...
Variables Affecting Phosphorescence and Fluorescence
Fluorescence and phosphorescence are essential phenomena in fields like analytical chemistry, biological imaging, and materials science, where they detect molecular properties and visualize cellular structures. Understanding the variables that influence these luminescent behaviors is crucial for maximizing accuracy and efficiency in their applications. These variables can broadly be grouped into chemical structure, solvent properties, and external conditions, each playing a distinct role in...
Determination of Crystal Structures
In the late 1800s, the revelation that light extended beyond visible wavelengths led to the discovery of X-rays by Wilhelm Roentgen. Recognized as high-energy electromagnetic radiation with short wavelengths, X-rays prompted exploration into their interaction with crystals. Max von Laue proposed in 1912 that the periodic arrangement of atoms, ions, or molecules in crystals would cause them to diffract X-rays, a hypothesis confirmed through experiments with copper sulfate and zinc sulfide...