Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
Assessment of liver viscoelasticity by using shear waves induced by ultrasound radiation force
Shigao Chen1, William Sanchez, Matthew R Callstrom
1Department of Biomedical Engineering, Mayo Clinic, 200 First St SW, Rochester, MN 55905, USA. chen.shigao@mayo.edu)
Purpose:
To investigate the value of viscosity measured with ultrasonographic (US) elastography in liver fibrosis staging and to determine whether the use of a viscoelastic model to estimate liver elasticity can improve its accuracy in fibrosis staging.
Materials And Methods:
The study, which was performed from February 2010 to March 2011, was compliant with HIPAA and approved by the institutional review board. Written informed consent was obtained from each subject. Ten healthy volunteers (eight women and two men aged 27-55 years) and 35 patients with liver disease (17 women and 18 men aged 19-74 years) were studied by using US elasticity measurements of the liver (within 6 months of liver biopsy). US data were analyzed with the shear wave dispersion ultrasound vibrometry (SDUV) method, in which elasticity and viscosity are measured by evaluating dispersion of shear wave propagation speed, as well as with the time-to-peak (TTP) method, where tissue viscosity was neglected and only elasticity was estimated from the effective shear wave speed. The hepatic fibrosis stage was assessed histologically by using the METAVIR scoring system. The correlation of elasticity and viscosity was assessed with the Pearson correlation coefficient. The performances of SDUV and TTP were evaluated with receiver operating characteristic (ROC) curve analysis.
Results:
The authors found significant correlations between elasticity and viscosity measured with SDUV (r = 0.80) and elasticity measured with SDUV and TTP (r = 0.94). The area under the ROC curve for differentiating between grade F0-F1 fibrosis and grade F2-F4 fibrosis was 0.98 for elasticity measured with SDUV, 0.86 for viscosity measured with SDUV, and 0.95 for elasticity measured with TTP.
Conclusion:
The results suggest that elasticity and viscosity measured between 95 Hz and 380 Hz by using SDUV are correlated and that elasticity measurements from SDUV and TTP showed substantially similar performance in liver fibrosis staging, although elasticity calculated from SDUV provided a better area under the ROC curve.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called a...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...