Bench comparative evaluation of a new generation and standard helmet for delivering non-invasive ventilation
C Olivieri1, R Costa, G Spinazzola
1Anesthesia and Intensive Care Medicine, Maggiore della Carità University Hospital, Novara, Italy.
Objective:
To evaluate the performance of a new helmet (NH) recently introduced into clinical use relative to that of the standard helmet (SH) in terms of delivering non-invasive continuous positive airway pressure (nCPAP) and pressure support ventilation (nPSV).
Design:
This was a bench study using a mannequin connected to an active lung simulator. The SH was fastened to the mannequin by armpit braces, which are not needed to secure the NH.
Measurements:
The inspiratory and expiratory variations in nCPAP delivered with two different simulated efforts (Pmus), were determined relative to the preset CPAP level. nPSV was applied at two simulated respiratory rates (RR) and two cycling-off flow thresholds. We measured inspiratory trigger delay (Delay trinsp), expiratory trigger delay (Delay trexp), time of synchrony (Time sync), trigger pressure drop (ΔP trigger), airway pressure-time product during the triggering phase (PTPt), the initial 200 ms from the onset of the ventilator pressurization (PTP 200), and the initial 300 and 500 ms from the onset of the simulated effort; this two latter parameters were expressed as the percentage of the area of ideal pressurization (PTP 300-index and PTP 500-index, respectively).
Results:
In nCPAP, at both Pmus, the differences between the two interfaces at both Pmus were small and clinically irrelevant. In nPSV, regardless of the setting, NH resulted in significantly smaller trigger delays, ΔP trigger, and PTPt. Time sync, PTP 200, PTP 300-index, and PTP 500-index were also significantly higher with the NH compared to the SH, irrespective of the setting.
Conclusions:
Compared to the SH, the NH is equally effective in delivering nCPAP and more effective in delivering nPSV, and it is used to avoid the need for armpit braces.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation II: ACLS Airway Management
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: