Anterior segment parameters measured with 2 optical devices compared to ultrasonic data
Gabor Nemeth1, Ziad Hassan, Eszter Szalai
1Department of Ophthalmology, Medical and Health Science Center, University of Debrecen, Debrecen, Hungary. nemeth222@yahoo.com
European Journal of Ophthalmology
|December 11, 2012
Summary
Optical devices and ultrasound show strong correlation for measuring central corneal thickness and anterior chamber depth. These findings support the interchangeability of these ophthalmic measurement techniques.
Area of Science:
- Ophthalmology
- Biomedical Optics
- Medical Imaging
Background:
- Accurate measurement of anterior segment parameters is crucial for ophthalmic diagnoses and surgical planning.
- Various optical and ultrasonic devices are available for these measurements, each with potential variations in accuracy and reliability.
Purpose of the Study:
- To compare anterior segment measurements obtained from two optical devices against an ultrasonic technique.
- To evaluate the consistency and correlation between different measurement modalities for axial length, anterior chamber depth, and central corneal thickness.
Main Methods:
- Axial length (AL) was measured using IOLMaster.
- Anterior chamber depth (ACD) was determined with IOLMaster, Visante optical coherence tomography (OCT), and Pentacam HR.
- Central corneal thickness (CCT) was measured using Pentacam HR and ultrasound.
Main Results:
- Strong correlations were found between Pentacam HR and ultrasonic measurements for CCT (r=0.89) and between Pentacam HR and Visante OCT for CCT (r=0.88).
- A significant correlation was observed between Pentacam HR and IOLMaster for ACD (r=0.95).
- Significant correlation was also found between AL and ACD measured with IOLMaster (r=0.67).
Conclusions:
- Optical devices and ultrasound demonstrate a strong correlation in measuring central corneal thickness.
- The anterior chamber depth measurements show significant correlation between Pentacam HR and IOLMaster, indicating potential interchangeability.
More Related Videos
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
Endoscopic Ultrasound (EUS):
Ultrasonography
Ultrasonography is an imaging technique that uses high-frequency sound waves to visualize the body's internal structures. It is a non-invasive and safe procedure that does not involve the use of ionizing radiation, making it widely used in various medical fields. Ultrasonography is used to study heart function, blood flow in the neck or extremities, certain conditions such as gallbladder disease, and fetal growth and development.
During an ultrasonography procedure, a handheld device called a...
During an ultrasonography procedure, a handheld device called a...
Imaging Studies II: Ultrasonography
IntroductionUltrasonography, or renal ultrasound, is a noninvasive medical imaging technique that uses high-frequency sound waves to visualize the kidneys, ureters, bladder, and surrounding tissues.Indications for Urinary System UltrasonographyUrinary system ultrasonography is indicated in various clinical scenarios, such as:Kidney Stones (Urolithiasis): To detect and monitor the size and presence of kidney or urinary tract stones.Hydronephrosis: To assess the dilation of the renal pelvis and...