Related Experiment Video
Updated: May 16, 2026

Synthesis and Calibration of Phosphorescent Nanoprobes for Oxygen Imaging in Biological Systems
Published on: March 3, 2010
Determination of total plasma hydroperoxides using a diphenyl-1-pyrenylphosphine fluorescent probe
Jonathan Santas1, Francesc Guardiola, Magda Rafecas
1Department of Nutrition and Food Science, XaRTA-INSA, University of Barcelona, 08028 Barcelona, Spain.
Abstract:
Plasma hydroperoxides (HPs) are widely accepted to be good indicators of oxidative stress. By means of the method proposed here, which uses diphenyl-1-pyrenylphosphine (DPPP) as a fluorescent probe, all types of plasma HP were determined. The limits of detection and quantification of the method were 0.08 and 0.25 nmol of cumene hydroperoxide (CHP) equivalents in 40 μl of plasma, respectively. The method is satisfactory in terms of precision (5.3% for 14.5 μM CHP eq., n=8), and the recoveries were 91% and 92% after standard additions of 26 and 52 μM CHP, respectively. The selectivity of the proposed method is higher than 96%. Moreover, optimization of the reaction conditions and the addition of ethylenediaminetetraacetic acid (EDTA) disodium salt and 2,6-di-tert-butyl-4-methylphenol (BHT) prevented the formation of HP artifacts during the analysis. Therefore, the proposed method is useful for simple and quantitative determination of total plasma HPs.