The effect of breathing irregularities on quantitative accuracy of respiratory gated PET∕CT
Boon-Keng Teo1, Babak Saboury, Reshma Munbodh
1Department of Radiation Oncology, University of Pennsylvania, Philadelphia, PA, USA. teok@uphs.upenn.edu
Purpose:
4D positron emission tomography and computed tomography (PET∕CT) can be used to reduce motion artifacts by correlating the raw PET data with the respiratory cycle. The accuracy of each PET phase is dependent on the reproducibility and consistency of the breathing cycle during acquisition. The objective of this study is to evaluate the impact of breathing amplitude and phase irregularities on the quantitative accuracy of 4D PET standardized uptake value (SUV) measurements. In addition, the magnitude of quantitative errors due to respiratory motion and partial volume error are compared.
Methods:
Phantom studies were performed using spheres filled with (18)F ranging from 9 to 47 mm in diameter with background activity. Motion was simulated using patient breathing data. The authors compared the accuracy of SUVs derived from gated PET (4 bins and 8 bins, phase-based) for ideal, average, and highly irregular breathing patterns.
Results:
Under ideal conditions, gated PET produced SUVs that were within (-5.4 ± 5.3)% of the static phantom measurements averaged across all sphere sizes. With breathing irregularities, the quantitative accuracy of gated PET decreased. Gated PET SUVs (best of 4 bins) were (-9.6 ± 13.0)% of the actual value for an average breather and decreased to (-17.1 ± 10.8)% for a highly irregular breather. Without gating, the differences in the SUV from actual value were (-28.5 ± 18.2)%, (-25.9 ± 14.4)%, and (-27.9 ± 18.2)% for the ideal, average, and highly irregular breather, respectively.
Conclusions:
Breathing irregularities reduce the quantitative accuracy of gated PET∕CT. Current gated PET techniques may underestimate the actual lesion SUV due to phase assignment errors. Evaluation of respiratory trace is necessary to assess accuracy of data binning and its effect on 4D PET SUVs.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...
Alterations in Respiration II
In Biot's breathing, the respiratory rate and depth are irregular, alternating between periods of deep gasping and apnea. Common causes include...
Respiratory Volumes and Capacities I
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...