Agreement between noninvasive oximetric values for oxygen saturation.

C Cahan1, M J Decker, P L Hoekje

  • 1Department of Medicine, Case Western Reserve University, Cleveland.

Chest
|April 1, 1990
PubMed
Summary

Pulse oximeters showed significant differences compared to the HP ear oximeter, especially in Black subjects and at lower oxygen saturation levels. These devices may not be interchangeable for accurate SaO2 measurements in clinical research.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Ventilatory and heart rate responses to hypoxia in pre- and post-menopausal ethnically Tibetan women residing at ∼3800 m in Mustang, Nepal.

The Journal of physiology·2026
Same author

Restoration of Sleep Architecture after Maxillomandibular Advancement: Success Beyond the Apnea-Hypopnea Index.

International journal of oral and maxillofacial surgery·2017
Same author

Respiratory and hematological adaptations of young and older Aymara men native to 3600M.

American journal of human biology : the official journal of the Human Biology Council·2017
Same author

Right ventricle dimensions and function in response to acute hypoxia in healthy human subjects.

Acta physiologica (Oxford, England)·2016
Same author

Scn1a dysfunction alters behavior but not the effect of stress on seizure response.

Genes, brain, and behavior·2015
Same author

Point/Counterpoint: a response to redefining chronic pain programs for the injured worker population.

Journal of back and musculoskeletal rehabilitation·2014

Area of Science:

  • Medical Devices
  • Clinical Measurement
  • Physiological Monitoring

Background:

  • The HP ear oximeter has been a standard for noninvasive oxygen saturation (SaO2) measurement.
  • Evaluating newer pulse oximeters for clinical research comparability is essential.

Purpose of the Study:

  • To assess the suitability of five pulse oximeters as replacements for the HP ear oximeter.
  • To determine the accuracy and agreement of pulse oximeters with the HP ear oximeter across different demographics and saturation levels.

Main Methods:

  • Isocapnic progressive hypoxia trials were conducted on 22 white and 6 black subjects.
  • Five pulse oximeters were compared against the HP ear oximeter for SaO2 measurements.
  • Statistical analysis involved comparing the differences between methods against their mean.

Related Experiment Videos

Main Results:

  • Overall, pulse oximeters showed a mean difference of 2.6 +/- 10.3% compared to the HP ear oximeter.
  • Differences were larger in Black subjects (5.1 +/- 9.2%) than in white subjects (1.9 +/- 10.2%).
  • Accuracy decreased significantly below 80% SaO2, with wider agreement limits (+/- 10%) above 85%.

Conclusions:

  • Pulse oximeters tend to overestimate SaO2 compared to the HP ear oximeter.
  • The observed discrepancies, particularly in Black individuals and at lower saturations, are clinically significant.
  • Current pulse oximeters are not interchangeable with the HP ear oximeter for SaO2 measurements in studies requiring high precision.