Noninvasive estimation of central blood pressure and the augmentation index in the seated position: a validation
Kozo Hirata1, Iwao Kojima, Shin-ichi Momomura
1Department of Cardiology, Saitama Medical Center, Jichi Medical University, Saitama, Japan. hirata@omiya.jichi.ac.jp
Objectives:
The aim of this study was to examine the utility of the generalized transfer function (GTF) and the late systolic shoulder of the radial pressure waveform (SBP2) for estimating seated central systolic blood pressure (c-SBP) and seated c-augmentation index (c-AIx).
Methods:
In 199 participants (182 cardiovascular disease patients and 17 normal individuals; mean age: 65.7 ± 10.6 years), c-SBP and c-AIx were estimated using the SphygmoCor and Omron HEM-9000AI systems. Both estimates were compared with carotid SBP and augmentation index (AIx). All measurements were performed with the participants in the seated position.
Results:
Mean c-SBP was 121.7 ± 18.6 mmHg and 122.6 ± 19.8 mmHg according to the GTF and SBP2 methods, respectively, and carotid SBP was 124.4 ± 19.2 mmHg. There were close correlations between the three values (GTF vs. carotid SBP: r = 0.986, SBP2 vs. carotid SBP: r = 0.975, GTF vs. SBP2: r = 0.989, all P < 0.0001), all of which met the AAMI SP10 criteria and the grade A British Society of Hypertension criteria. The relationship between c-AIx estimated using the GTF method and carotid AIx was comparable to that reported previously for supine individuals (r = 0.703, P < 0.0001). The seated carotid-radial artery amplifications of the first and second harmonics were identical to those reported in previous supine studies. Both methods produced reproducible estimates of c-SBP and c-AIx (r = 0.997-0.910, all P < 0.0001).
Conclusion:
The present study indicates that the GTF and SBP2 methods can be used to estimate c-SBP and c-AIx in seated individuals.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessing Blood pressure in the Leg
Preparation: