Gas chromatographic retention of alkyl phosphates on ionic liquid stationary phases
Brandon M Weber1, James J Harynuk
1Department of Chemistry, University of Alberta, Edmonton, Alberta, Canada T6C 2X4.
Abstract:
Retention behaviors of alkyl phosophates were studied on a series of ionic liquid gas chromatography columns. The selectivity of the IL columns for alkyl phosphates were compared with a 5% phenyl column as a route to evaluating the potential use of IL columns in the analysis of alkyl phosphates in petroleum samples in both one- and multi-dimensional GC. Most interestingly, we demonstrate for the first time the dependence of elution order on separation temperature for members of a homologous series of compounds. At low temperatures it was found that trihexyl phosphate eluted before trioctyl phosphate, while at higher temperatures this pattern was reversed.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Ion-Exchange Chromatography
Analyte Adsorption and Distribution