Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
New approaches based on modified Gaussian models for the prediction of chromatographic peaks
J J Baeza-Baeza1, C Ortiz-Bolsico, M C García-Álvarez-Coque
1Departament de Química Analítica, Universitat de València, c/Dr. Moliner 50, 46100 Burjassot, Valencia, Spain. Juan.Baeza@uv.es
Abstract:
The description of skewed chromatographic peaks has been discussed extensively and many functions have been proposed. Among these, the Polynomially Modified Gaussian (PMG) models interpret the deviations from ideality as a change in the standard deviation with time. This approach has shown a high accuracy in the fitting to tailing and fronting peaks. However, it has the drawback of the uncontrolled growth of the predicted signal outside the elution region, which departs from the experimental baseline. To solve this problem, the Parabolic-Lorentzian Modified Gaussian (PLMG) model was developed. This combines a parabola that describes the variance change in the peak region, and a Lorentzian function that decreases the variance growth out of the peak region. The PLMG model has, however, the drawback of its high flexibility that makes the optimisation process difficult when the initial values of the model parameters are far from the optimal ones. Based on the fitting of experimental peaks of diverse origin and asymmetry degree, several semiempirical approaches that make use of the halfwidths at 60.65% and 10% peak height are here reported, which allow the use of the PLMG model for prediction purposes with small errors (below 2-3%). The incorporation of several restrictions in the algorithm avoids the indeterminations that arise frequently with this model, when applied to highly skewed peaks.
More Related Videos
10:25Construction of Models for Nondestructive Prediction of Ingredient Contents in Blueberries by Near-infrared Spectroscopy Based on HPLC Measurements
Published on: June 28, 2016
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Chromatographic Methods: Terminology
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...